Draw 3 ml ACD-A for APC 20 in 20 ml syringe. Draw 8 ml ACD-A for APC 60 in 60 ml syringe. |
Transfer 1 ml ACD-A from the 20 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 20. Transfer 2 ml ACD-A from the 60 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 60. |
|
Draw venous blood using ACD-A syringe for APC 20 to the 3/4 oz mark (approx. 22 ml). Draw 60 ml venous blood using ACD-A syringe for APC 60. |
Transfer total syringe volume into blood chamber of PD (red port). |
|
Load SmartPReP® with PD, close the lid, press the green start button. NOTE: To prevent system imbalance, use correct balance weight when processing only one PD.
|
After processing, use syringe with spacer and withdraw PPP (white port) until air bubbles are present. Transfer contents to yellow cup on surgical field. |
|
Using syringe without the spacer, resuspend Platelets with remaining PPP. Transfer contents to red cup on surgical field.
|
Mix APC+ activator. Transfer contents to clear cup on surgical field. |