Quick Set-up - SmartPRep Office


Step 1 of APC+ ProcessStep 1

Draw 3 ml ACD-A for APC 20 in 20 ml syringe.

Draw 8 ml ACD-A for APC 60 in 60 ml syringe.

Step 3 of APC+ ProcessStep 2

Transfer 1 ml ACD-A from the 20 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 20.

Transfer 2 ml ACD-A from the 60 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 60.

Step 2 of APC+ ProcessStep 3

Draw venous blood using ACD-A syringe for APC 20 to the 3/4 oz mark (approx. 22 ml).

Draw 60 ml venous blood using ACD-A syringe for APC 60.

Step 4 of APC+ ProcessStep 4

Transfer total syringe volume into blood chamber of PD (red port).

Step 5 of APC+ ProcessStep 5

Load SmartPReP® with PD, close the lid, press the green start button.

NOTE: To prevent system imbalance, use correct balance weight when processing only one PD.

 

Step 6 of APC+ ProcessStep 6

After processing, use syringe with spacer and withdraw PPP (white port) until air bubbles are present.

Transfer contents to yellow cup on surgical field.

Step 7 of APC+ ProcessStep 7

Using syringe without the spacer, resuspend Platelets with remaining PPP.

Transfer contents to red cup on surgical field.

 

Step 8 of APC+ ProcessStep 8

Mix APC+™ activator.

Transfer contents to clear cup on surgical field.

 